Oncentra/Monaco – Pencil beam and Collapsed

Cone Convolution

Patient tissue distribution is represented by a 3D density matrix where the

properties in each voxel are either derived from pixel values in a CT image

or from user specified values of mass or electron density.

It is assumed that the specified density values correspond to tissue material

of “standard” compositions (see ICRP 23 and ICRU 44).

The Hounsfield number assigned to each standard tissue is mapped from

the calculated electron density using relations given by Knöös et al [38].

…for charged particle equilibrium conditions CC calculates dose to medium

while PK calculates dose to a large water cavity in medium. Both algorithms

consider attenuation of the beam as per medium

thus;

Pencil kernel and for point kernel

Dublin 2018

15

,

,

Previously TMS-Helax, Oncentra MasterPlan