Table of Contents Table of Contents
Previous Page  219 / 879 Next Page
Information
Show Menu
Previous Page 219 / 879 Next Page
Page Background

Process Chart

To Do list – Brain Met. Stereo RT

Is a part of a treatment procedure

Made by a Physicist who is involved in treating the

patient

The Physician signs the document ‘on-line’

And participates in the session

Mandatory while doing any Stereo RT at our center

(brain, lung or bone met.)

Each action is performed in a specific order

(‘

modus operandi’

)

Patient installation

Table positioning

Control imaging

Treatment arcs delivery

Medical validation

Patient out of dressing room (

hh.mm)

Done (Y/N)

Place treatment beams and positionning beams in

correct order

Select CB-CT

Place a radio-opaque marker on carbon fiber head

support

Install patient

Carve up mask above contralateral eyebrows.

Place mask et radio-opaque marker on patient

Perform shifts and markings

Place a radio-opaque marker on mask at isocenter

Acquire CBCT and perform 3D-registration of table

shifts 3D

Acquire kV image ‘Ant Head-AP’ (270°-Arm)

Record distances of points to graduation axis

Acquire KV image ‘Post Head-AP’ (90°-Arm)

Record distances of points to graduation axis

Deliver first transverse Arc Table at 0°

Select KV-270° (90°-Arm)