Table of Contents Table of Contents
Previous Page  379 / 879 Next Page
Information
Show Menu
Previous Page 379 / 879 Next Page
Page Background

98

99

100

101

0

50

100

150

200

250

Moyenneparpatientdu%depointsayantungamma<

1

Date

ORL P1 IUPESM Nouvelles limites APRES chgt algo

95

96

97

98

99

100

0

50

100

150

200

250

Moyenne parpatient du%de points ayantun gamma < 1

Date

ORL P1 IUPESM Nouvelles limites APRES chgt algo

0,0

0,4

0,8

1,2

1,6

2,0

2,4

2,8

0

50

100

150

200

250

ORL P1 IUPESMNouvelles limites APRES chgt algo

N

°

contrôle

pecial cause: TPS : changement of algo. version

42

Performance indicators

Control charts

Process under control

Stable ou superior

In daily practice at ICL

:

< 5% points out of limits

Process under control and

statistically capable

S

pecial cause identified

Performance indicators

dispersion

mean

-

Head-and-neck treatments -

5. IMRT QC results (portal dosimetry) monitored with SPC

S

TANDARDIZE

Moving-Range (%)

Individual value. (%)

EWMA (%)

New limits