Value driven tolerance limits – RP162

Criteria for Acceptability of Medical Radiological Equipment used in Diagnostic Radiology, Nuclear Medicine

and Radiotherapy. RP 162 (European Commission), 2013

R

ADIOTHERAPY

Physical Parameter

Suspension

Level

Reference

(IEC (2007,

2008c)

clause

numbers

unless

stated)

Type

Symmetry of electron fields (max/min

ratio)

>1.05

A

Maximum ratio of absorbed dose

(max/min ratio)

1.09 See IEC

A

Dose monitoring system

7

Weekly calibration check

>2 %

A

Reproducibility

>0.5 %

A

Proportionality

>2 %

A

Dependence on angular position of gantry and

beam limiting device

>3 %

A

Dependence on gantry rotation

>2 % - electron

radiation

>3 % - X-radiation

A

Stability throughout the day

>2 %

A

Stability in moving beam radiotherapy

See IEC

A

Depth dose characteristics

8

X-radiation

Penetrative quality

>3 % or 3 mm.

A

Depth dose and profiles

>2 %

IPEM (1999) B