Summary: measurements in small static fields

• Consensus so far on current good practice for reference and relative dosimetry in

static small MV photon fields:

Careful experimental setup

Choice of suitably small detector which is known to minimally perturb fluence

Approximately correct for volume averaging and energy dependence of

detector

Corroboration of data

Current research efforts in small field dosimetry focus on the determination of

detector specific output correction factors.

Detectors requiring output corrections greater that 5% are not recommended

for dose determination in small fields.

The IAEA TECDOC will include a consistent set of such data and will be an

international code of practice for small static field dosimetry.