DETECTORS FOR MEASUREMENT-2D/3D detectors

The range of dose levels used for calibration must cover the measurement

range.

6 films per point dose recommended for calibration

Films should be assigned doses in a random pattern (avoid trending to

inhomogeneities)

Film handling tips

25

Utretch 2016

Use the same scanning protocol

for calibration and for measuring

Need to be scanned in the same orientation [orientation of “needle like mictro-

crystals (EBT)]

For best results, scan one day following irradiation (min. two hours)

Take into account

corrections for the scanner

. Linearity and non homogeneity.