Page Background

1st Commandment

Explained

Metric for γ-analysis

o

“Overreliance on the insensitive metric is counterproductive to quality

improvement and can lead to the sense of complacency among the

clinical physicists.”

44

o

“IMRT and VMAT commissioning, along with product validation, would

benefit from the retirement of the 3%/3 mm passing rates as a primary

metric of performance, and the adoption instead of tighter tolerances,

more diligent diagnostics, and more thorough analysis.”

Med Phys 2013 – Nelms et al