Dose to tissue?
• D
m,m
is the ‘natural’ quantity from MC
• To report D
w
then if convert back to
water – uncertainty in s
w,m
• Makes little difference for soft tissues
• CT numbers converted using ICRU or
ICRP but who is ‘standard?
• D
w
clinical experience (Ma et al?)
• Dosimetry protocols
• NTCP, TCP reported in dose to water
• Uncertainties lower
• Target is a ‘water’ cell in bone
• Report both?
External Beam
For
Against
• Mass energy absorption coeffs quite different between tissue
types and water
• D
m,m
and D
w,m
can be very different
• Accurate dosimetry requires atomic composition
• MC-D
w,m
good substitute for MC-D
n.m
for all energies and
tissue types studied
(Enger et al PMB 2012)
Brachytherapy