OF: small fields
Parallel to beam axis
Perpendicular to beam axis
Organic scintillator
(W1 Standard Imaging)
0.0%
1.0%
2.0%
3.0%
4.0%
5.0%
Relative difference normalized to EFD OF
Exradin W1 Cal. A
Exradin W1 Cal. B
Exradin W1 Cal. C
The calibration methodology proposed by
the manufacturer is not appropriate for
measuring output factors for small fields.
Calibration using smaller fields and the
detector axis parallel to beam axis should
-5.0%
-4.0%
-3.0%
-2.0%
-1.0%
0.0 1.0 2.0 3.0 4.0 5.0 6.0 7.0 8.0 9.0 10.0 11.0
Side of equivalent square field size at isocenter (cm)
Pinpoint
Relative differences of OF normalized to EFD-3G OF.
OF are referred to a 10 x 10 cm
2
field size.
be used.
These results show that the CRL and gain
depend on field size and orientation of the
scintillator.