Previous Page  870-871 / 2239 Next Page
Information
Show Menu
Previous Page 870-871 / 2239 Next Page
Page Background

1.025

1.030

1.035

1.040

1.045

1.050

Relative mass stopping power ratio

60Co

4 MV

6 MV

10 MV

15 MV

18 MV

Variation of

s

water,air

for ion chamber

measurements

10

7

photons

Outside beam

Inside beam

0.990

0.995

1.000

1.005

1.010

1.015

1.020

0.0 1.0 2.0 3.0 4.0 5.0 6.0 7.0 8.0 9.0 10.0 11.0 12.0

Off axis distance (cm)

Stopping power ratios along a radii at 10 cm depth in water

Field limit

Constant spectra over the field

Calculations performed with sprrznrc

From T. Kn

öos