Repeatability

54

Variation arising when all efforts are made to keep conditions constant by using

55

the same instrument and operator, and repeating during a short time period.

56

Expressed as the repeatability standard deviation (SD

r

); or % repeatability

57

relative standard deviation (%RSD

r

).

58

59

Reproducibility

60

The standard deviation or relative standard deviation calculated from among-

61

laboratory data. Expressed as the reproducibility relative standard deviation

62

(SD

R

); or % reproducibility relative standard deviation (% RSD

R

).

63

64

65

4.

Method Performance Requirements:

66

See Table 1.

67

68

Table 1. Method Performance requirements

a

69

Analytical range

1–50

b

Limit of Quantitation (LOQ)

≤ 1

b

Recovery

90-110%

Repeatability (RSD

r

)

8%

Reproducibility (RSD

R

)

15%

a

Concentrations apply to: a) ‘ready-to-feed” liquids “as is”; b) re-

constituted powders (25 g into 200 g of water); and c) liquid

concentrates diluted 1:1 by weight.

b

μg /100 g reconstituted final product

70

71

5.

System suitability tests and/or analytical quality control:

72

Suitable methods will include blank check samples, and check standards at the

73

lowest point and midrange point of the analytical range. Methods must be capable of

74

resolving α-carotene from lycopene and

β

-carotene.

75

76

6.

Reference Material(s)

: Neither NIST nor JRC produce a certified reference material for

77

α-carotene in infant formula.

78

79

7.

Validation Guidance

:

80

Recommended level of validation:

Official Methods of Analysis

SM

.

81

82

8.

Maximum Time-To-Result

: No maximum time.

83

84