Background Image
Table of Contents Table of Contents
Previous Page  1291 / 1456 Next Page
Information
Show Menu
Previous Page 1291 / 1456 Next Page
Page Background

MU range checking

In house NKI development

Plans following a certain protocol, e.g. prostate

– Amount of MU for a VMAT plan will be similar for each patient

– Depends a little on patient size, etc

MU range check

– If patient does not fall within the range, something may have gone wrong

– Check by physics

– About 5-10%

– Usually anatomical reasons

– Some errors found (wrong dose specification point)