Table of Contents Table of Contents
Previous Page  132 / 154 Next Page
Information
Show Menu
Previous Page 132 / 154 Next Page
Page Background

ISO/WD

© ISO 2016 – All rights reserved

xxii

samples with a high chloride content, weigh a smaller test portion e.g. 5 g of

reconstituted or RTF product

.

7.3

Transfer 2,5 ml precipitating solution I (4.3) and 2,5 ml precipitating

solution II (4.4) into the tube. Complete to 50 ml with water. Mix well. If foam

impacts the constant volume, one or two drops of defoaming agent (3.12)

shall be used.

7.4

Centrifuge at 12 000

g

for 5 min at 4°C. Then equilibrate to room

temperature.

7.5

Accurately transfer either 10 ml supernatant from steps 7.2 to 7.4 or

weigh an appropriate aliquot of RTF or reconstituted powder (e.g. 25 g

accurate to 0,1 mg). For samples with a high chloride content, weigh a smaller

test portion e.g. 5 g of reconstituted or RTF product.

Into a 120 ml sample beaker or autosampler cup. Add 5 ml nitric acid solution (4.5), and 50 ml of

water before titration. Add a magnetic stirring rod (if the titrator does not have a build-in rod

stirrer). Place the autosampler cup or beaker onto a magnetic stirrer and stir until dissolved or

finely suspended.

7.6

The pH of the test solution shall be below 1,5. In case of any doubts,

check by means of a pH-meter and, if necessary, add a little more nitric acid

solution (4.5).

8

Instrument operating conditions

8.1

Check and maintenance of the combined silver electrode

Rinse electrode with deionized water and wipe before use. Renew the electrolyte 1 mol/l KNO

3

(4.9) periodically per manufacturer’s recommendations. If fat sticks to the electrodes during a

series of analyses, then eliminate it by briefly immersing the electrode in acetone. The silver

electrode shall be stored in 1 mol/l KNO

3

(4.9) or per manufacturer’s recommendations after

appropriate cleaning.

NOTE: In place of the combined silver electrode, separate silver and reference electrodes may also

be used.

8.2

Titration

Connect the combined silver electrode to the titration apparatus, according to the manufacturer's

indications. Ensure that the titration vessels are correctly placed on the autosampler and there are

enough reagents: both nitric acid solution (if added automatically) (4.5) and 0,1 mol/l AgNO

3

(3.7

or

[J-CTS5

]

4.1). If no autosampler is available, then place the sample solutions manually under the

titration equipment.

Put the wash solution (4.6) in the washing position if an auto sampler used. Ensure that the volume

of wash solution is adequate.

Under continuous stirring and without touching the electrode, titrate the sample solution

automatically with 0,1 mol/l standardized silver nitrate solution (3.7 or

[J-CTS

6]

4.1), up to the end

potential. The consumption of 0,1 mol/l of silver nitrate solution should be recorded automatically

and can be read from the titrator software, or documented in the titrator operating records. For

manual titration, using a burette add 0,1 mol/l standardized silver nitrate solution, until the end

potential has nearly been met. Then continue to titrate slowly until the end point is met as

observed by the two small (about 0,05 ml) silver nitrate solution.

Special case: determination of very low amounts of chloride

2016.03 (APRIL 2017) - MTE-01/03

MLT REPORT

FOR ERP USE ONLY

DO NOT DISTRIBUTE