Repeatability

53

Variation arising when all efforts are made to keep conditions constant by using

54

the same instrument and operator, and repeating during a short time period.

55

Expressed as the repeatability standard deviation (SD

r

); or % repeatability

56

relative standard deviation (%RSD

r

).

57

58

Reproducibility

59

The standard deviation or relative standard deviation calculated from among-

60

laboratory data. Expressed as the reproducibility relative standard deviation

61

(SD

R

); or % reproducibility relative standard deviation (% RSD

R

).

62

63

64

4.

Method Performance Requirements:

65

See Table 1.

66

67

Table 1. Method Performance requirements

a

68

Analytical range

1–50

b

Limit of Quantitation (LOQ)

≤ 1

b

Recovery

90-110%

Repeatability (RSD

r

)

8%

Reproducibility (RSD

R

)

15%

a

Concentrations apply to: a) ‘ready-to-feed” liquids “as is”; b) re-

constituted powders (25 g into 200 g of water); and c) liquid

concentrates diluted 1:1 by weight.

b

μg /100 g reconstituted final product

69

70

5.

System suitability tests and/or analytical quality control:

71

Suitable methods will include blank check samples, and check standards at the

72

lowest point and midrange point of the analytical range. Methods must be capable of

73

resolving lycopene from α-carotene and

β

-carotene.

74

75

6.

Reference Material(s)

:

76

SRM 1869. Please contact Dr. Melissa Phillips, Research Chemist, NIST for materials at

77

melissa.phillips@nist.gov

or (301) 975-4134.

78

79

7.

Validation Guidance

:

80

Recommended level of validation:

Official Methods of Analysis

SM

.

81

82

8.

Maximum Time-To-Result

: No maximum time.

83