Delivery uncertainties - IMRT

H&N

Prostate

Dosimetric effects of delivery

uncertainties in

dose rate

,

gantry angle

, and

MLC leaf

how intensity modulation is achieved?

sliding window

larger number of

overlapping segments and MU

RapidArc

rotation

larger

apertures and less MU

2%-2 mm

g

for systematic MLC leaf bank shifts

of 0.5, 1, 1.5, and 2 mm. Points that failed in red.

Betzel GT et al, Med Phys 39(10) 2012

positions

for RapidArc vs sliding

window IMRT RT plans

The susceptibility to MLC leaf position

errors is inversely proportional to the

mean leaf gap width

Geometric errors in smaller apertures

larger impact on dosimetric errors.