Previous Page  2448-2449 / 3107 Next Page
Information
Show Menu
Previous Page 2448-2449 / 3107 Next Page
Page Background

Patient Specific Pretreatment QA

1.

IMRT or VMAT plan is approved by Radiation Oncologist

2.

Redundant dose calculation

Single point or full 3D dose distribution

Based on treatment planning output or machine control settings

3.

Pretreatment QA

1.

Plan is copied to measurement device

2.

Dose is recalculated on measurement device

Convert all fields to 0-gantry, 0-collimator, 0-table rotation

(analyze results on a field-by-field basis)

3.

Export calculated dose (at the plan of the detector)