IMRT – the small field challenge

Small fields - Inhomogeneities - electronic disequilibrium

effects

Fraction of scatter becomes significant for small fields

Small no. of MU’s (step-and shoot !)

STATUS: OK, proves the stable relationship between the dose monitor setting and

the linac CAX output in static circumstances at 0° gantry angle. The exact curve

fitting methodology should be included in the acceptance procedure.

-5

0

5

10

15

20

1

10

100

1000

deviation from linear fit

deviation of absorbed dose [%]

MU

500MU/min

400MU/min

200MU/min

100MU/min

Figure 3: Deviation from linearity of the output in function of the amount of delivered MUʼs.

UZ Brussel 17/12/2009 v4.0

Dose-MU linearity

Minimum number of MU at RMH is 3MU

VN Hansen et al,Phys. “

Quality assurance of the dose

delivered by small radiation Segments”

Med. Biol.

43 (1998) 2665–2675