69

è

Conventional RT

: 10

1

parameters/plan

è

IMRT

:

10

3

parameters/plan

è

TomoTherapy

:

10

4

parameters/plan

Complex treatments and the need for patient specific QA

Impossible to verify all

combinations with machine

oriented QA experiments

In IMRT/VMAT/Tomotherapy the classic entrance dose in vivo, which was a proved

technique to detect large errors, became difficult,

and yet it are these complex

techniques which really require patient specific QA

Even advanced TPS contains

only limited model

of the beam and machine …”

Is it possible to cover all operation parameter combinations

in machine related QA for

advanced IMRT delivery techniques? Maybe we should

focus on the end result

the

delivered dose to the patient and

relate any dosimetric discrepancies between intended and

delivered dose to the impact on treatment effectiveness

?!

Prove the plan works with an experimental verification

1) Dose, MU calc

2) Information transfer TPS->R&V

3) Dose delivery