no inhomogeneity correction (CT/MRI/UH only in the background for

catheter/applicator reconstruction)

- assumption of full scattering condition (infinite medium)

-

no correction for source movement of HDR source (only dwell times)

- no correction for applicator/catheter absorption

- no correction for seed absorption

Limitations of TG-43 formalism

Model-based dose calculation algorithms (MBDCAs

) either explicitly simulate the transport

of radiation in the actual media or employ multiple dimensional scatter integration techniques

to account for the dependence of scatter dose on the 3D geometry.

Beaulieu et al. Med Phys 39.6208-36. 2012