Technical Considerations
•
QA
Long-term Energy Stability (321 procedures over a 4.5-year period)
Percentage of D
50
measurements
within +/- 2 mm and +/- 1mm from expected value
4 MeV 6 MeV 9 MeV 12 MeV
[-2mm, 2 mm] 99,0 100,0 98,1
96,9
[-1 mm, 1 mm] 88,9 91,6 82,6
90,0