Dosimetrical comparison between multilumen MammoSite
and Contura Multilumen Balloon applicators
(median values in % of prescribed dose)
V95 V150 V200 Skin
max
Rib
max
ML MammoSite (8 pts.)
98.0
31.6
7.4
92.5 92.7
Contura (12 pts.)
97.6
28.2
7.5
98.2 104.0
Both applicators
have shown the ability of dose shaping for providing
sufficient target coverage while concurrently limiting dose to skin and rib
Fu et al. University of Pittsburgh Cancer Institute, Pittsburgh, PA