Example for HDR

192

Ir BT source for breast balloon applicator

Category

Optimum level

Assumptions

Source strength

2% PSDL traceable calibrations.

Treatment planning

3% Reference data with the appropriate bin width.

Medium dosimetric corrections

3% Balloon filled with standard level of contrast

agent, no consideration or composition of

chestwall, lung, or breast.

Scatter dosimetric corrections

7% A non-scalar correction for skin dose is

needed, and will require an advanced dose

calculation formalism to properly account for

radiation scatter conditions in the patient.

Dose delivery including registration

of applicator geometry to anatomy 7% Accurate QA concept for commissioning and

constancy checks, especially for source

positioning and applicator/ source path

geometry, appropriate imaging techniques

(either small slice thickness, 3D sequences or

combination of different slice orientations),

applicator characterization.