De Brabandere, et al.,

Brachytherapy

12, 580-588 (2013)

Seed Reconstruction Uncertainties: CT & MRI

MRI-based seed reconstruction

was less accurate than CT, with a

mean interobserver variation in

seed positioning of 3 mm (1 SD).

This resulted in a non-negligible

mean interobserver variation in

D

90

of about 7% for T1 + T2.