Dosimetry using Dosimetry using a Grid-Based

Boltzmann equation Solver (Acuros)

GBBS algorithms used primarily for neutron transport and

shielding problems

Method evaluated for brachy in the literature as early as 2000

The first GBBS algorithm incorporated in a commercially available

192

Ir brachytherapy TPS was Acuros

Acuros is based on the Attila GBBS developed at Los Alamos

National Laboratory, optimized for brachytherapy and later also for

external beam therapy (XB) and even CT scatter correction (CTS).

Method can be reviewed in a number of publications:

Daskalov GM et al 2000 Med. Phys. 27(10):2307

Daskalov GM et al, 2002 Med. Phys. 29(2): 113

Gifford K A et al, 2006 Phys. Med. Biol., 51(9): 2253

Gifford et al 2008 Med. Phys. 35(6): 2279

+

BV-Acuros user manual