Absolute Calibration
Reference condition
:
(this can differ from institute to institute)
10*10 cm field at Isocentre
100cm SSD delivers
1 Gy (100cGy) per 100 MU at d
max
Basic Dose Caculation
Principles 2012
5
Other reference conditions:
10*10 cm field at isocentre
100cm- d
max
SSD delivers
1Gy per 100 MU at d
max
10*10 cm field at isocentre
90cm SSD delivers
1Gy per 100 MU at d
10cm