SMPR Key Points

1.

Method Performance Requirements

:

Table 1:  Analytical Range & LOQ Based on Matrix 

Parameter 

Finished Products 

Raw Materials 

Analytical range ppm* 

0.5 – 338 

1,250 ‐ 12,500 

Limit of Quantitation ppm* 

0.4 

1,000 

Table 2: Method Performance Requirements as a Function of Range

Parameter 

Ranges (µg/g)* 

< 10 ‐ 15 

>15 ‐ 50 

>50 – 500 

>500 – 4,000 

>4000 – 12,500 

Recovery (%) 

80 ‐ 110 

90 ‐ 107 

95 – 105 

95 – 105 

97 – 103 

% R t bilit

 epea a y 

(RSD

r

% Reproducibility 

(RSD

R

12 

10 

* Measured as individual forms of Vitamin D and pre‐Vitamin D 

Comments Submitted (if any)

One Comment Submitted

“The two certified reference materials I 

mentioned were vitamin D2 and vitamin D3.  

We have other vitamin D standards in the 

attached but they are the metabolites.”

Cerilliant Vitamin D2V‐024 Vitamin D3V‐025

1 mg/ml CRM in Ethanol

St k h ld D dd t SMPR

a e o ers:  o we a o 

 as a 

requirement to use?