”On-board / In-room imaging” for BT:

Imaging of source progression

Real-time flouroscopy of

source

Flouroscopy during 2031

treatment sessions in 370

patients

2/2031 errors detected

(0.1%)

Affected 2/370 (0.5%) of pts

Swapped guide tubes

Wrong applicator length

Nose et al, Int J Radiation Oncol Biol Phys, Vol. 97, No. 4, pp. 858-865, 2017