Evaluate the movement:

Head

à

minimize the compliance of mask, neck support &

bite

block

Eye :

Involuntary:

drift (ex: 0.5 -1 mm/sec)

oscillations

(ex: amp 0.5mm, freq 2-3 Hz)

Voluntary or fainted :

«how long is the treatment ?? »

(1cm, 0.1 sec)

II. Adopt the strategy :

Positioning with Xrays & clips

à

set-up with tenths of mm

Transfert referential to video : monitoring without X rays

Gating (manual or automatic) with video & Coaching

Back to X rays if necessary. If mouvement ++: sucker/lens, margins or abandon

III. Requirements/Consequences :

- Orthogonal X rays, panels, infrared camera, electronic pen or screen,…

-

Dose rate

high enough to make a short session (15sec -1 min),

low enough to minimize doses if need to interrupt

make your calculations… : 30 Gy/min; tot dose 4x15= 60Gy

1 sec target off

à

0.5 Gy ( ~ 1 %), ~1 mm