Protec9ve!shielding!

Collimator!channels!

Leksell

®!

Coordinate!

Frame!

Pa9ent!!

posi9oning!system

!

Radia9on!sources!

Isocenter/!

Target!in!the!brain!

Frame-based Stereotactic Radiosurgery

Positioning Accuracy

Accuracy!and!stability!of!posi9oning!in!radiosurgery:!!

long!–term!results!of!the!Gamma!Knife!system.!

!

!

!

!

!

!

!

!

!!!!!Heck!B!et!al!

Med!Phys!2007!Apr;!34(4):!1487J95!

Graf2Chromic2films2

densitometric!measurements!

!

!

!X:!J!0.014+/J!0.09mm!

!

!

!Y:!!!!0.013+/J!0.09mm!

!

!

!Z:!J!0.002+/J!0.06mm!

!

!

All!measured!data!were!within!a!sphere!of!!

0.2mm

!radius!

!

2

MRI>based2target2defini0on2

!

!

!X:!0.06+/J0.09mm!

!

!

!Y:!0.04+/J0.09mm!