SMPR Key Points

Concentrates

Parameter

Requirement 

THC, THCA, CBDA, CBD, 

Individually Reported

Requirement 

(Additional, 

Desirable 

Cannabinoids 

(Table 1b), 

including CBN)

Limit of 

Quantitation (LOQ)  

(% w/w)

≤ 0.3

≤ 0.3

Analytical Range 

(% w/w)

≤ 0.3 – ca. 100

≤ 0.3 – ca. 50

Dried Plant Material

Parameters

Ranges (% w/w)

≤ 0.3 – 1

> 1 ‐ 10 > 10 – ca. 

100

Recovery (% w/w)

95 – 105

97 ‐ 103 98 ‐ 102

% RSD

r

≤ 5

≤ 4

≤ 2

% RSD

R

≤ 7

≤ 5

≤ 3

Parameters

Ranges (%)

(Pertains to Cannabinoids in Table 1b)

0.1 – 1

> 1 ‐ 25 > 25 – ca. 50

Recovery (%)

95 – 105

97 ‐ 103

98 ‐ 102

% RSD

r

≤ 5

≤ 4

≤ 2

% RSD

R

≤ 7

≤ 5

≤ 3

Parameter

Requirement

(Pertains to Cannabinoids 

in Table 1b)

Limit of Quantitation (LOQ)  (%)

≤ 0.1

Analytical Range (%)

0.1 – ca. 50*

*Lower concentrations may be acceptable as applicable for 

cannabinoids listed in Table 1B.

Comments Submitted

5 Comments were received for SMPR –dried 

materials.

1. The proposed procedures should add (a) 

Instrument Detection Limit (IDL), and (b) 

Method Detection Limit (MDL)

Action: WG decided to use LOD which is the 

equivalent of IDL, which uses blank matrix 

samples;  LOD uses matrix samples.  MDL was 

purposefully removed by the WG.