Dose calculation – AAPM-TG43

G(r,θ): The geometry function at radial distance r and polar angle θ. It is

a dimensionless quantity.

From a physics point of view the geometry function simply provides an

effective inverse square-law correction based upon the spatial

distribution of radioactivity within the source core.

It depends only on the 3D volume shape of the active core of the source

and not on the radionuclide, nor on the encapsulation!

It neglects scattering and attenuation!