![Show Menu](styles/mobile-menu.png)
![Page Background](./../common/page-substrates/page0552.jpg)
Uncertainties and Errors
Systematic errors
– Have the same influence in every brachytherapy fraction
• Errors in application commissioning process
• MR distorsions
• Incorrect reconstruction procedure
• Magnification errors
Random errors
– Uncertainties that differ between brachytherapy sessions
• Visibility of the applicator
• Accuracy of image fusion
• MR distorsions related to patient specific interaction with
magnetic field