Table of Contents Table of Contents
Previous Page  107 / 1082 Next Page
Information
Show Menu
Previous Page 107 / 1082 Next Page
Page Background

S94

ESTRO 36 2017

_______________________________________________________________________________________________

interest, only 0.2% of voxels differ by >5%, showing good

agreement throughout. Results from measured delivery

errors, such as those in figure 1, will also be presented.

Figure 1: Dosimetric comparison between planned and

delivered doses for a HDR brachytherapy treatment in a

phantom with an introduced error.

Conclusion

Real-time dosimetric treatment verification is possible

with our source tracking system combined with MaxiCalc.

Fast dose calculation based on measured source dwell

positions is achieved and overcomes the limitation of

current TPSs.

References

1.

Smith, R L., et al. Medical physics 43.5 (2016):

2435-2442.

2.

Daskalov, G M., et al. Medical physics 25.11

(1998): 2200-2208.

PV-0187 Source dwell time and transit time

measurement for a HDR afterloading unit

T.L. Chiu

1

, B. Yang

1

, H. Geng

1

, W.W. Lam

1

, C.W. Kong

1

,

K.Y. Cheung

1

, S.K. Yu

1

1

Hong Kong Sanatorium & Hospital, Medical Physics &

Research Department, Happy Valley, Hong Kong SAR

China

Purpose or Objective

To evaluate dwell time and transit time of HDR

brachytherapy treatment by an in-house fluorescent

screen based QA system. Since dosimetric effect would be

directly affected by source dwell time, an accurate QA

method on temporal accuracy is essential.

Material and Methods

The system included a fluorescent screen (Kodak, Lanex

regular screen) which converts the radiation signal to

optical signal and a high-speed camera with frame rate up

to 500 fps and pixel resolution of 1280X720. The temporal

resolution was 2 ms. A catheter in which an Ir-192 source

would be loaded was fixed on the fluorescent screen and

the camera was placed 30 cm away from the screen. The

whole system was light-shielded. When the source

travelled inside the catheter, the camera would capture

images on the fluorescent screen sequentially. Source

position was traced out by locating the centroid of the

captured image. The accuracy of dwell time was assessed

by measuring 3 different dwell times, namely, 1 s, 0.5 s &

0.1 s. According to a white paper from vendor, transit time

for separations below 35 mm would occupied part of the

next dwell time and those for separations above 35 mm

would have 0.1 s compensation. Thus, the influence of

transit time on dwell time was studied by measuring 0.5 s

dwell time under 3 different dwell separations, namely, 6

cm, 4 cm & 0.5 cm. Dwell time was assessed by counting

the number of images in which source positions were

unchanged to the subsequent image. Transit time was the

time between two dwell positions.

Results

Fig. 1 demonstrated the capability of this QA system by

showing a source transit process and Fig. 2 indicated that

measured dwell time was affected by source separation.

Table 1(a) tabulated the measured dwell time for 3

different assigned dwell times with 5 mm separation

between source dwell positions. In all three scenarios, the

dwell time at starting position was close to the assigned

value. Dwell time at next dwell position experienced a

larger discrepancy up to 40% for 0.1 s dwell time. This

discrepancy in dwell time was due to the transit time for

which control computer could not fully account. Hence,

dwell time would be shorter than the assigned value

except at the starting position. Table 1(b) tabulated

measured dwell times at 3 different source separations

with 0.5 s assigned dwell time to assess the compensation

method stated. Discrepancy could be up to 0.33 s in 6 cm

separation. Transit time occupied a larger portion of the

dwell time for longer source separation.

Conclusion

Dwell time and transit time could be measured using the

fluorescent QA system with uncertainty down to 2 ms.

High temporal resolution in this system helped measure

the transit time accurately which could hardly be achieved

in commonly used QA systems. The effect of transit time

on actual source dwell time could be significant and was

not fully accounted for by treatment computer. Clinically

possible combinations, like 0.5 s dwell time and 5 mm

separation, could have a dosimetric error of 8%.

PV-0188 Improved class solutions for prostate

brachytherapy planning via evolutionary machine

learning

S.C. Maree

1

, P.A.N. Bosman

2

, Y. Niatsetski

3

, C. Koedood

er

1

, N. Van Wieringen

1

, A. Bel

1

, B.R. Pieters

1

, T.

Alderliesten

1

1

A cademic Medical Center, Radiation oncology,

Amsterdam, The Netherlands

2

Centrum Wiskunde & Informatica, Amsterdam, The