Table of Contents Table of Contents
Previous Page  851 / 1082 Next Page
Information
Show Menu
Previous Page 851 / 1082 Next Page
Page Background

S835

ESTRO 36 2017

_______________________________________________________________________________________________

In each case the CI is improved using the VMAT solution.

The average CI for SCF plans was 0.64, this increased to

0.81 when using VMAT. The GI appeared to increase in

several cases, with improvement only shown in 4 out of 10

cases. In each case, NTOF was reduced to close to zero.

Perhaps surprisingly for a VMAT solution, the 5Gy volume

was reduced for 9 out of 10 cases. The average 5Gy volume

for the SCF plans was 36.65cc; this reduced to 26.55cc

when using VMAT.

Conclusion

Paddick CI has proven to be a reliable metric for

conformity, while NTOF gives a good indication of how

baggy the prescription isodose is. However, Paddick GI was

not found to be a reliable indicator of low dose spread,

especially for small volumes. In many cases the 40% and

80% volumes had each been reduced and conformity

improved but a higher GI was recorded.

The Eclipse VMAT solution is adequate for a wide range of

tumour shapes, sizes and locations and can be used for all

single-target SRS patients.

EP-1570 The dynamic jaw mode of tomotherapy:

Better neural structure protection for advanced NPC

patients?

P.T. Fang

1

, T.Y. Lu

1

, M.N. Wu

2

, M.Y. Huang

1

, C.J. Huang

1

1

Kaohsiung Medical University Hospital, Radiation

Oncology, Kaohsiung, Taiwan

2

Kaohsiung Medical University Hospital, Neurology,

Kaohsiung, Taiwan

Purpose or Objective

This study investigated the neural structure protection

effects of the dynamic jaw mode of tomotherapy for

advanced nasopharyngeal cancer (NPC) by comparing use

of the dynamic jaw mode and the fixed jaw mode with

different field widths.

Material and Methods

Twenty patients with locally advanced NPC were selected.

All T classifications were T3-4. Plans were using the

simultaneous integrated boost technique in 3 dose levels.

Four plans were generated utilizing the 2.5 cm dynamic,

2.5 cm fixed, 5.0 cm dynamic, and 5.0 cm fixed jaw

modes. Plan efficiency was evaluated in terms of monitor

unit and beam-on-time measurements. Plan quality was

evaluated using homogeneity index (HI) and conformity

index (CI). Dose indices of neural structures such as the

optic pathways, temporal lobe, and hippocampus, as well

as of other organs at risk (OAR), were compared.

Volumetric parameters of the hippocampus and temporal

lobe were also assessed. Only plans of the same field width

were compared with one another.

Results

The comparison of the 2.5 cm dynamic jaw plan with the

2.5 cm fixed jaw plan indicated no differences between

plan efficiency, HI and CI results, or the dose indices of

OARs within the radiation field. However, use of the 2.5

cm dynamic jaw mode significantly improved the

maximum dose to the left optic nerve (p=0.005) and

chiasm (p=0.035). It also lowered the maximum dose and

mean dose to the hippocampus (both p values were 0.035).

In addition, the volume receiving 5 Gy (V5) values for the

temporal lobe and hippocampus were significantly smaller

when the dynamic jaw mode was used (p=0.035; p=0.013).

The comparison of the 5.0 cm dynamic jaw plan with the

5.0 cm fixed jaw plan also indicated no differences

between plan efficiency, HI and CI results, or the dose

indices of OARs within the radiation field. However, use of

the 5.0 cm dynamic jaw mode significantly improved the

maximum dose to the left and right optic nerve (p<0.0005)

and chiasm (p=0.005). The mean dose to the temporal lobe

was significantly improved using the dynamic jaw mode

(p=0.013). It also lowered the maximum dose and mean

dose to the hippocampus (both p=0.035). Finally, the

dynamic jaw mode also significantly reduced the V5, V10,

and V20 values for the temporal lobe and hippocampus.

Conclusion

When the same field width, the dynamic jaw mode

provided better neural structure protection than the fixed

jaw mode, in addition to reducing the low-dose volume to

the temporal lobe and hippocampus. These advantages

were more pronounced when using a larger field width.

There were no significant differences, however, in plan

quality and efficiency between the two modes for the

same field width. The above results indicate that for the

treatment of locally advanced NPC patients, use of the 2.5

cm dynamic jaw mode rather than the 2.5 cm fixed jaw

mode currently used in clinical practice would provide

better neural structure protection and lower low-dose

volumes to the temporal lobe and hippocampus with equal

plan

efficiency.

EP-1571 Radiotherapy treatments using a prototype

MLC design

P. Nitsch

1

, Y. Li

1

, T. Netherton

1

, P. Balter

1

, S. Gao

1

, M.

Muruganandham

1

, S. Shaitelman

2

, S. Frank

2

, S. Hahn

2

, A.

Klopp

2

, L. Court

1

1

The University of Texas MD Anderson Cancer Center,

Radiation Physics, Houston, USA

2

The University of Texas MD Anderson Cancer Center,

Radiation Oncology, Houston, USA

Purpose or Objective

We are evaluating the clinical efficacy of a prototype

multi-leaf collimator (MLC) design which obviates the

need for collimating jaws. The new MLCs are 1.0cm wide,

potentially giving increased reliability, and have a

maximum speed of 5.0cm/sec. The increased leaf width

may reduce the achievable intensity modulation, but the

impact of this may be mitigated by the increase in MLC

speed and by using IMRT/VMAT treatment planning. Here

we evaluate (1) whether clinically acceptable plans can be

created using such an MLC design, and (2) the agreement

between the planned and delivered dose distributions.

Material and Methods

IMRT, VMAT, field-in-field and electronic compensator

plans were created in the Eclipse treatment planning

system using the prototype MLC design and a flattening-

filter-free 6MV beam, for the following treatment sites:

head/neck, lung (standard fractionation, palliative and

SBRT plans), cervix (pelvis and extended fields), intact

breast (left and right), prostate (SBRT and involved

nodes), and whole brain treatments. The planned dose

distributions and DVHs reviewed for clinical acceptability

by radiation oncologists, and compared with our original

clinical plans (120leaf Millennium MLC, 0.5cm MLCs).

Delivered dose distributions for IMRT and VMAT plans were

evaluated using the ArcCHECK array.

Results

In most situations, the plan quality (particularly

homogeneity in the target) was highest for IMRT (with

multiple collimator angles), followed by VMAT, electronic

compensator, and field-in-field. The optimal IMRT and

VMAT plans were typically considered clinically

acceptable, while the electronic compensator and field-