Software upgrades: QA software

QA phantom (Delta 4) for patient pre-treatment verification

Software upgrade: No specifics given..

After upgrade, new calibration method needed

Ion chamber output modelled differently for small and large

fields

Everything seemed ok, but..

After some time, doing a statistical process analysis (trend

analysis), on average lower pass rates were found.

Still working on how to handle this issue…