![Show Menu](styles/mobile-menu.png)
![Page Background](./../common/page-substrates/page0362.jpg)
Table assisted rotation correction
Gevaert (and verellen) IJROBP 83, 2012 p467:
Using Brainlab mask system, 40 pats
Before and after IGRT on Novalis couch:
Mean 3D:
Before: M=1.91 mm
±
1.25 mm and
after:
M=0.58 mm
±
0.42 mm.
Mean rotational errors:
Before:
-0.10
±
1.03 (vert), 0.23
±
0.82 (long) and -0.09
±
0.72 (lat)
After:
0.01
±
0.35 (vert),0.03
±
0.31 (long) and 0.03
±
0.33 (lat)
(intrafraction, after approx 15 min)
A ≥0.5
°
rotation was identified as threshold for coverage loss. (Volume covered by
prescription isodose would have decreased by 5% in this population)
Ohtakara R&O 102, 2012 p198: Brainlab vs standard mask:
Both are suitable for 6DOF brain SBRT set-up, with standard mask requiring 0.5 mm
larger margin