Department of Medical Physics

Department of Medic l Physics

Dwells

DosePoints

Vaginal Cylinder Treatment planning

Example: VagCyl.Ø30mm,VL=10cm TL=35cm

, 2.5 mm step-size

Active source pos.

Inactive

Lat. dose points on applicator surface

Lat. dose points in 5mm tissue depth

:

158%

±

11%

: 100%

±

6%

2.5mm step-size

Apex DP [%]

0mm 5mm

354,49 101,32

Lateral DP [%]

0mm 5mm

143,98 92,66

161,30 101,82

168,74 105,76

165,98 104,04

147,87 94,40

Active Dwell

Pos time [s]

1 14,20

2 14,60

3 14,92

4 14,91

5 14,95

6 14,55

7 14,91

8 14,81

9 15,62

10 16,19

11 17,05

12 17,69

13 18,09

Ʃ202,5s