Performance Characteristics

Analytical Range:  0.2 to 4.2 mg/L in measured solutions but could not be 

determined per AOAC.  LOQ:  0.2 mg/L  Accuracy / Recovery:  Recovery was 97%. 

Could not be ascertained in the absence of cis‐isomer standards.  Precision:  RDS was 

0.23%.  Reproducibility:  RSD <0.92%

System Suitability

Conditions not provided to allow suitability to be replicated.

—All instrument configuration; instrument details, including serial numbers and 

acquisition information; and column details, including identification, were properly 

recorded during laboratory operations. Equipment calibration, including signal 

description and baseline noise determination, was performed before tests were 

executed.

Do you recommend this method be 

adopted as a First Action and 

published in the Official Methods 

of Analysis of AOAC 

INTERNATIONAL?  Please specify 

rationale.  

This is the basis of a functional method. Several things could improve and expand 

the method. It may be more applicable to measuring free lutein and zeaxanthin than 

the esters. I would recommend that some things be adjusted before recommending 

first action status.  Sample prep to address additional products. Assessment of the 

composite‐specific absorbance. Inclusion of standards and beta‐cryptoxanthin.

Recommendation for the Method