Previous Page  981 / 1363 Next Page
Information
Show Menu
Previous Page 981 / 1363 Next Page
Page Background

Requirement

Recommendation/Comment

System

1.5 T or 3.0 T wide bore (70 cm) scanner

Set-up

Flat table & immobilisation;

utilise high channel RF coil arrays & intensity correction

Distortion

3D gradient correction; measure residual.

High BW; Optimise shim where needed;

B0 mapping still not widely available

Protocol

2D TSE & DIXON (±C) primary planning sequences;

3D important in specific cases (e.g. T2 MR-only, T1 SRS)

Functional & motion assessment not standard

Resolution/

coverage

maximum 1mm or 1 pixel fat-water shift

2-3/0 mm slices (2D) standard; 4 mm functional (e.g. DWI)

Single concatenation

3D isotropic as required

30 cm scan coverage

MR-Simulator