SPSFAM Heavy Metals ERP Book
(grams), M
DIW = mass of the DIWused to dilute the juice concentrate
Table 2016.04E. Recommended Data Analysis Method a settings for m/z 75 Data point sampling: 1
Start threshold: 0.3 Stop threshold: 0.5 Peak location: Top
Smoothing: enabled
Detection filtering: 5 point
Baseline reset (No. of points): >10 If leading or trailing edge: <50 Baseline preference: Drop else tangent
skim b Peak area, counts: >2000 All other parameters should be left at default values a Integration parameter values are specific to the MassHunter Data Analysis software. b Specific terminology used by MassHunter Data Analysis software for dropping a baseline from the start to the stop of the peak rather than a tangent line.
© 2016 AOAC INTERNATIONAL