Vi-CELL BLU -Application Notes
Results
Cell Concentrations for 8 serial dilutions of CHO cells Low Concentration ~0.5M -5M Range. 36 samples per dilution
6.00
R = 0.9996
5.00
4.00
3.00
Cell Concentration (M/mL)
2.00
1.00
0.00
0%
10%
20%
30%
40%
50%
60%
70%
80%
90%
100%
Dilution %
Figure 3. CHO Cells Low Concentration Range Dilution Series Data
Cell Concentrations for 7 serial dilutions of CHO cells High Concentration ~2.6M-13M Range. 30 samples per dilution
14.00
R = 0.9988
12.00
10.00
8.00
6.00
Cell Concentration (M/mL)
4.00
2.00
0.00
0%
10%
20%
30%
40%
50%
60%
70%
80%
90%
100%
Dilution (%)
Figure 4. CHO Cells High Concentration Range Data
Additional sample plates were run using a smaller range of dilutions to confirm the performance of the instruments. These were repeated 3 times within a 16 hour period using the same stock supply of cells run on the same instrument. During this period cell population increase is considered minimal and the source material for analysis effectively the same.
Characterized by Ingenuity | 4