Why switch to MC dose computation?

PB algorithm (especially in combination with range

shifter) inaccurate for two reasons:

Source: RS5 reference manual, RSL

Nuclear halo effect

Each pencil beam is modelled by 2 Gaussians (MCS,

nuclear halo)

Lack of handling nuclear halo properly within the

range shifter, then transporting the beam through

vacuum (instead of air) and large heterogeneities

(patient surface): causes lack of modelling accuracy

especially for low energies where a greater angular

spread of the protons is expected.

Lateral heterogeneities

Each spot is split into 19 sub-pencil beams.

In case of large spot sizes (combination of range

shifter and larger gaps) the distance between subspots

becomes larger than anatomic density variations within

the patient.

Courtesy N. Schreuder, Provision Knoxville, 2017