AOAC First Action Method 2016.02 (Biotin) MLT Report

Page 10 of 16

Figure 2: MLT Sample 8B (00859RF00 / PZGP859)

4. RESULTS AND DISCUSSION

4.1. The overview

The details of the 12 laboratories from 10 different countries participated in the MLT study is given in Table

2. The laboratories were requested to analyse biotin and biocytin in the samples, however none of the

samples were found to have biocytin and only biotin results were reported and evaluated in the study. All

the biotin results presented in the report are expressed as

µg/100g

of sample as received basis from AOAC

International.

Three of the eleven laboratories (Labs 10, 11 and 12) could only report the results of practice samples on

the due date of the study, 27 Jan 2017. In general, the results from Lab 7 is showing low bias which was

investigated further. The possible reasons for the low bias could be the practice followed in sample

preparation especially the steps related to loading of immunoaffinity column and the elution technique

where triple backflush is specified. The investigation from Lab 7 reported that the analysts did not follow the

backflush technique effectively and that may have caused the incomplete elution of biotin from the

cartridges. The laboratory also reported that they had deviated from the protocol in reconstitution of powder

samples by taking 10g of powder in 80mL. The results from Labs 7, 10, 11 and 12 are included in the MLT

report and for statistical evaluation of the results to calculate precision of the analytical method.

2016.02 (FEBRUARY 2017) BIO-02 MLT REPORT