Laboratory & Quality Services

P. Stuyvesantweg 1

8937 AC LEEUWARDEN

Postbus 226

8901 MA LEEUWARDEN

Pagina

8

van

8

Overview

Table 7. Overview performances

Analyte

Mn

Cu

Fe

Zn

Mg

P

Na

Ca

K

Lower analytical

range

a

0,003

0,010

0,2

0,01

0,09

0,04

0,5

0,4

6

Upper analytical

range

b

5

5

125

25

375

1250

625

1875

1875

Limit of

Quantitation

(LOQ)

0,003

0,010

0,2

0,01

0,09

0,04

0,5

0,4

6

Repeatibility ≤

(RSD)

c

1,2%

0,5%

1,3%

1,8%

1,5%

1,0%

0,7%

1,7%

0,8%

Recovery

d

97,4%

100,3%

96,9% 96,0% 97,6%

103,7%

95,5%

98,9%

98,5%

Reproducibility ≤

(RSD)

e

1,4%

1,7%

0,9%

1,0%

1,2%

1,0%

1,2%

1,2%

1,4%

a

Lower analytical range is equivalent to the LOQ.

b

Upper analytical range is equivalent to the highest standard (5) of the calibration curve used on daily

basis calculated by multiplying this standard by a dilution factor of 25.

c

Repeatibility of NIST SRM 1849a calculated by using the residues of the duplicates on 9 different days.

d

Overall average for 3 independent days using NIST SRM 1849a and the laboratory control sample

CM01/07.

e

Reproducibility of NIST SRM 1849a calculated by using the average results of the duplicates on 9

different days.

*mg/100g reconstituted final product.

MTE-03-SLV