LOD and LOQ:

Table 2. Determination of LOD and LOQ by measuring 24 blanks

calculated according to appendix L: "AOAC Recommended

Guidelines.."

Instrument

Lower

lineair limit

Sample

PLOQ

Analyte

LOD

mg/100g

a

LOQ

mg/100g

a

PLOQ

mg/100g

b

mg/100g

c

Ca

0,1

0,4

5

125

Cu

0,002

0,010

0,002

0,05

Fe

0,09

0,2

0,05

1,25

K

2,6

6

0,6

15

Mg

0,03

0,09

1

25

Mn

0,001

0,003

0,002

0,05

Na

0,2

0,5

0,2

5

P

0,01

0,04

0,5

12,5

Zn

0,004

0,01

0,01

0,25

a

Calculated by multiplying the LOD and LOQ by a dilution factor 25.

b

PLOQ is equivalent to the lower limit of the calibration curve.

c

Calculated by multiplying the PLOQ by a dilution factor of 25.