Background Image
Table of Contents Table of Contents
Previous Page  196 / 620 Next Page
Information
Show Menu
Previous Page 196 / 620 Next Page
Page Background

Laboratory & Quality Services

P. Stuyvesantweg 1

8937 AC LEEUWARDEN

Postbus 226

8901 MA LEEUWARDEN

Pagina

2

van

8

Linearity:

Table 1. Measurement of standard solutions to determine linearity for Ca, Cu, Fe, Mg, Mn, Na, P and Zn

using six calibration standards.

Analyte

Conc.

mg/kg

a

Recovery,

%

b

RSD,

%

c

Analyte

Conc.

mg/kg

a

Recovery,

%

b

RSD,

%

c

Analyte

Conc.

mg/kg

a

Recovery,

%

b

RSD,

%

c

Ca

50

100.6

0.4

K

6

102.1

2.2

Na

2

101.8

1.8

100

100.8

0.3

30

102.7

1.8

10

101.5

1.1

200

100.7

0.3

150

101.2

1.2

50

100.8

0.8

500

98.3

0.9

300

99.1

1.1

100

98.7

1.2

750

99.9

0.4

750

100.7

2.0

250

99.6

0.8

975

100.4

0.3

975

99.6

1.6

325

101.1

1.4

Cu

0.02

96.7

0.8

Mg

10

100.9

1.3

P

5

101.4

0.8

0.1

101.3

1.0

20

100.8

0.9

25

101.3

1.6

0.5

100.9

0.6

40

100.4

0.3

125

100.0

1.4

1

97.9

1.3

100

98.4

0.4

250

97.8

1.4

2

99.7

0.4

150

100.0

0.6

500

100.5

1.0

2.6

100.8

1.0

195

100.3

0.3

650

101.8

1.2

Fe

0.5

101.8

2.4

Mn

0.02

101.0

0.9

Zn

0.1

101.1

1.7

2.5

101.8

1.5

0.1

100.8

0.8

0.5

101.9

1.1

12.5

101.6

1.6

0.5

100.7

0.6

2.5

101.6

0.8

25

98.2

1.8

1

97.5

1.4

5

98.1

1.3

50

100.6

1.7

2

99.9

0.4

10

100.7

1.3

65

101.2

1.8

2.6

100.8

0.5

13

100.9

0.6

a

Expected concentration

b

Overall average for 3 independent days

c

Precision for 3 separate independent days

MTE-03-SLV

FOR ERP USE ONLY

DO NOT DISTRIBUTE