HPLC Determination of Total Tryptophan in Infant Formula and

Adult/Pediatric Nutritional Formula Following Enzymatic

Hydrolysis

Abbott Nutrition, Division of Abbott Laboratories

page

2

b)

HPLC system with autosampler, column Oven to maintain 30°C, Fluorescence

Detector, binary or quaternary HPLC Pump with degasser, HPLC peak area

integrator or chromatography data processor.

c) Constant temperature oven capable of maintaining 50 (+/- 1) degree Centigrade

2. Instrument Parameters and Conditions

a)

Mobile Phase Flow Rate : 0.5 mL /minute

b)

Column Temperature: 30 degrees Centigrade

c)

Injection volume: 10

L.

d)

Detector settings: Excitation wavelength to 295 nm and emission wavelength to

345 nm

e)

Typical PMT setting: 7 for Agilent G1321A (see F.3.a)

f)

Run time: 7.5 minutes, or as needed to fully collect the 5-methyl-DL-tryptophan

peak.

E.

REAGENTS

1.

Standards

L-Tryptophan USP Standard Cat #1700501, or equivalent.

2.

Chemicals

a)

Methanol, HPLC grade

b)

Sodium Phosphate Monobasic, HPLC grade.

c)

Phosphoric Acid, (85%)

d)

Trizma Base

e)

Hydrochloric Acid, 6N.

f)

Laboratory Water – minimum of 18 mega ohm conductivity

g)

5-Methyl-DL-tryptophan Sigma Cat # M-0534, or equivalent.

h)

Protease E (enzyme) Sigma Cat # P-5147, or equivalent.

3.

Solutions

a)

Hydrochloric Acid, 1.0N Solution.

b)

Trizma Buffer, 0.1M, pH 8.5.