Determination of Choline in Infant Formula,

Adult/Pediatric Nutritional Formula, food and

feed

ESI LC-MS/MS

(applicable to the determination of free and total choline in

Food/feed, infant formula, adult nutritionals and nutritional

supplement.)

A. Principle

Choline and esters of this are extracted by methanol and

water. Free choline is determined on the extract. Total

choline is measured after alkaline hydrolysis at elevated

temperature of the extract, thereby converting the bound

forms in to free choline. The resulting free choline is

quantified by isotope dilution LC-MSMS.

B. Apparatus

(

a

) Column. - Synergi 2,5μ polar-RP 100 x 3.00 mm

(Phenomenex), or equivalent.

(

b

) Liquid chromatograph. - Thermo Dionex Ultimate 3000

RS 600bar HPLC system, or equivalent.

(

c

) Detector. - AB Sciex Q-Trap 3200 mass spectrometer, or

equivalent.

(

d

) Data system. – Analyst 1.5.1 and Dionex

Chromatography MS link 2.10.0.2906

(

e

) Balance.—Readable to at least 0.0001 g (Satorius, or

equivalent).

(

f

) Balance.—Readable to at least 0.001 g (Kern, or

equivalent).

(

g

) Incubator. – Capable of 37 ± 5°C

(

h

) Centrifuge.

(

i

) Shaking table. (Inside Incubator)

(

j

) Oven. - Capable of 102 ± 2°C

(

k

) Centrifuge tubes.—plastic , 50 mL with screw cap.

(

l

) Extraction vessels. – conical flasks, 250 ml with stoppers.

(

m

) Digestion vessels. Conical flasks, 500 ml with stoppers.

(

n

) Volumetric flasks. 50, 100 and 500 ml and 1 L

(

o

) Syringe filters.— 0.45 μm.

(

p

) Pipets.—5000 μL adjustable

(

q

) Pipet tips.

C. Reagents

(

a

) Purified Water, USP; ≥18 Mohm/cm.

(

b

) Methanol. – HPLC-grade (for sample preparation)

(

c

) Methanol. – LC-MS grade (for LC-MS/MS mobile phases

preparation)

(

d

) Ammonium formate. LC-MS grade, crystalline, ≥99.0%

(

e

) Formic acid. LC-MS grade, ≥99.9%

(

f

) Potasium hydroxide. – p.a., crystaline, ≥85%

(

g

) Hydrochloric acid. –p.a., 37%

D. Standards

(

b

) Choline bitartrate reference standard. – U.S.

Pharmacopeia (Rockville, MD; Cat. No. 1133536).

(

d

) Choline-1,1,2,2-d4 bromide. - C/D/N Isotopes (Quebec,

Canada; Cat. No. D-1600).Purified.

E. Solutions Preparation

(

a

) Ammonium formate buffer. – 100mM, pH 3.0

Quantitatively transfer 1.12 ±0.01 g ammonium formate into

a 1 L beaker add 800 ml water, dissolve, adjust pH to 3,0

with concentrated formic acid, transfer to volumetric flask

and fill to volume with water.

(

b

) Mobile phase. –

(

1

) Mobile phase A. – Measure 200 mL ammonium formate

buffer into a 1 L volumetric flask and dilute to the mark with

water.

(

2

) Mobile phase B. – Methanol LC-MS grade

(

c

) Potassium hydroxide solution. – 2 M Dissolve 56.1 g

reagent grade KOH pellets in water in a 500 mL volumetric fl

ask, and dilute to volume with water. Storage: plastic bottle,

room temperature. Expiration: 1 year.

(

d

) Hydrochloric acid.- 1.6 M. Measure 67.5 mL

concentrated HCl into 500 mL volumetric flask and dilute to

the mark with water. Standardization is not required.

Expiration: 1 year.

(

e

) Choline bitartrate solution (approximately 4 mg/g choline

ion). – Dry ca. 1.5 g choline bitartrate at 102 °C to constant

weight. Accurately weigh 975 mg dry choline bitartrate as

quickly as possible (because of hygroscopicity), and transfer

to a 100 mL flask. Dissolve choline bitartrate in water, and

dilute to volume with water and accurately weigh the

solution, the concentration is calculated on mass to mass

basis.

(

f

) Choline bitartrate intermediate solution, (approximately

400 µg/g choline ion). — Accurately weigh 10 g Choline

bitartrate solution into a 100 mL flask. Dilute to volume with

water and accurately weigh the solution.

(

g

) Choline Stock solution, (approximately 12 µg/g choline

ion). — Accurately weigh 15 g Choline bitartrate

intermediate solution into a 500 mL flask. Dilute to volume

with water and accurately weigh the solution. Dispense the

solution into plastic centrifuge tubes, each receiving ≥10 mL.

Storage: freezer, –20°C. Expiration: 1 year.

(

h

) Choline-1,1,2,2-d

4

bromide Stock solution,

(approximately 80 µg/g). – Accurately weigh 70 mg Choline-

1,1,2,2-d

4

bromide as quickly as possible, and transfer to a

500 mL flask. Dissolve choline-1,1,2,2-d

4

bromide in water,

and dilute to volume with water and accurately weigh the

solution. Dispense the solution into plastic centrifuge tubes,

each receiving ≥15 mL. Storage: freezer, –20°C. Expiration:

1 year.

(

i

) ISTD standard solution (approximately 2.3 µg/g choline-

1,1,2,2-d

4

). – Accurately weigh 14 g Choline Stock solution

into a 200 mL flask. Add 4 mL of 1.6 M Hydrochloric acid

and 3 ml of 2 M Potassium hydroxide solution. Dilute to

volume with water and accurately weigh the solution. This

amount of internal standard is sufficient to 6 standards and 18

samples.

(

j

) Standard solution (approximately 1.2 µg/g of L- choline).

– Accurately weigh 10 g Choline Stock solution into a 100

mL flask. Dilute to volume with water and accurately weigh

the solution.

(

k

) Working standards. — Prepare fresh on day of analysis.

(

1

) Level 1 calibration standard. - Accurately weigh 1 g

standard solution and 4 g ISTD standard solution into a 100

mL flask. Dilute to volume with water and accurately weigh

the solution. Cap and mix well. Transfer to an autosampler

vial and cap.

(

2

) Level 2 calibration standard. - Accurately weigh 3 g

standard solution and 2 g ISTD standard solution into a 50

mL flask. Dilute to volume with water and accurately weigh

the solution. Cap and mix well. Transfer to an autosampler

vial and cap.

(

3

) Level 3 calibration standard. - Accurately weigh 10 g

standard solution and 2 g ISTD standard solution into a 50

mL flask. Dilute to volume with water and accurately weigh

the solution. Cap and mix well. Transfer to an autosampler

Carn-06 (Choline)