Evaluation / Reporting
Level 1
Minimum
standard for reporting
Level 2
Advanced
standard for reporting
Level 3
Research
oriented reporting
Dose delivery pattern
HDR
Absorbed Dose (Gy)
Fraction size (Gy)
Number of Fractions
Interval between fr. (h)
PDR
Absorbed Dose (Gy)
Pulse size (Gy)
Number of pulses, and
interval between pulses
Number of fractions and
interval between fr.
LDR
Absorbed Dose (Gy)
Dose Rate (Gy/h)
Number of Fractions
and Treatment-time
Biological dose reporting
EQD2 values for traget and OAR (endpoints)
Respective
α/β
values* for the target and OAR
Respective T1/2 of recovery*
Applied recovery model, mono- or bi-exponential
*For the moment - advise to follow GEC-ESTRO recommendations
α/β
=3Gy for late
Effects in OAR and 10 Gy for tumour response, and T
1/2
of 1.5h for both.
EUD or other biologically effective dose model calc. for target and OAR endpoints
TCP or NTCP calculation, with the model explicitly stated
Source and Dose calculation (
Radionuclide, Source model, Source Strength, Dose calculation algorithm
)
Total treatment-time (TT) in hours or days of EBRT, BT and overall TT