FOR EXPERT REVIEW PANEL USE ONLY

NOT FOR DISTRIBUTION

2

Data are well organized and clearly presented

Method Clarity:

Method is clearly written.

Method lists all the necessary supplies and analysis steps

Method Safety Concerns:

No major safety concerns are noted

Pros/Strengths:

Uses established methodology and commonly available equipment

Fast chromatographic analysis

Method describes alternative eluants and reagents preparations procedures – from

commercially available kits as well as “from scratch” using common chemicals

Data for all SPIFAN matrices are presented

Linearity and other parameters are evaluated and compared with SMPR

Cons/Weaknesses

Accuracy data and intermediate reproducibility data for many amino acids don’t meet SMPR

Method is not suitable for analysis if Tryptophan

Supporting Data

-

Method Optimization:

Some method optimization may be needed if baseline separation for amino acids is not

achieved using chromatographic conditions described in the method

Performance Characteristics:

Analytical Range and Linearity: